6.1006.520 Metrosep A Supp 5 - 150  
Metrosep A Supp 5 - 150    
Housing material   PEEK  
Carrier material   Polyvinyl alcohol with quarternary ammonium groups  
Capacity   70 (µmol Cl)  
Particle size (µm)   5 µm  
Eluents  

Carbonate eluent (standard eluent)

Sodium hydrogen carbonate: 1.0 mmol/L ;168 mg/2 L
Sodium carbonate: 3.2 mmol/L ; 678 mg/2 L

Hydroxide eluent, modified

Sodium hydroxide: 120.0 mmol/L ; 600 mg/2 L
p-Cyanophenol: 4.0 mmol/L ; 476 mg/2 L

Chromate eluent

Sodium hydrogen carbonate: 4.0 mmol/L ; 168 mg/2 L
Sodium carbonate: 12.8 mmol/L ; 2712 mg/2 L

 
Standard flow   0.7 mL/min  
Organic modifier   0 - 100 %  
Max. pressure   15 MPa  
Maximum flow   0.8 mL/min  
pH range   3 - 12  
Storage   In the eluent  
Column type   Column  
Column dimensions (mm)   150 x 4.0  
Regeneration  

Contamination by low-valency hydrophilic ions:

a) Rinse with ultrapure water (25 min at 0.3 mL/min)
b) Rinse with 10-fold concentrated eluent (100 min at 0.3 mL/min)
c) Rinse with ultrapure water (25 min at 0.3 mL/min)
d) Rinse with eluent (100 min at 0.3 mL/min)

Contamination by high-valency hydrophilic ions:

a) Rinse with ultrapure water (25 min at 0.3 mL/min)
b) Rinse with 5% acetonitrile (20 min at 0.3 mL/min)
c) Rinse with 100% acetonitrile (60 min at 0.3 mL/min)
d) Rinse with 50% acetonitrile (10 min at 0.3 mL/min)
e) Rinse with ultrapure water (50 min at 0.3 mL/min)
f) Rinse with eluent (100 min at 0.3 mL/min)

System peak shifted (regeneration method with column oven):

a) Rinse with concentrated eluent 1 mol/L (25 min at 0.4 mL/min)
b) Keep at 45 ... 50 °C for 10 ... 12 hours (no rinsing)
c) Rinse with standard eluent (minimum 40 min at 0.4 mL/min)

 
Application range  

Applications

  • standard anions
  • F-, Cl-, Br-, I-
  • ClO2-, ClO3-, ClO4-, BrO3-
  • ClO4-
  • BrO3-
  • Cr (VI) (CrO42-)
  • method development
  • difficult matrices
  • difficult separating problems
 
  Technical documents