In this way it is possible to determine anions and cations even in cutting oil emulsions, dairy products or body fluids such as blood and urine.
«ProfIC 7» also offers the possibility of automatic inline dilution of the sample. This takes place at an external position of the 858 Professional Sample Processor. In combination with a transfer tubing, the 800 Dosino carries out the high-precision liquid handling steps.
The intelligent MagIC Net™ software controls both the dialysis and the dilution steps. MagIC Net™ provides for all calculations; only the dilution factor needs to be entered in the sample table. In addition, the software provides for intelligent dilution, i.e. dilution takes place automatically depending on the predefined concentration limits. The software monitors the results and makes logical decisions.
The anion system is equipped with sequential suppression. Very high sensitivity is achieved by this unique combination of chemical and downstream CO2 suppression. The flexibility of the carbonate eluent is complemented by the low background conductivity of a hydroxide eluent.
The 858 Professional Sample Processor with Pump carries out the sample transfer. Numerous sample racks for volumes 0.5...500 mL are available.