Contamination by low-valency hydrophilic ions:
a) Rinse with ultrapure water (25 min at 0.3 mL/min) b) Rinse with 10-fold concentrated eluent (100 min at 0.3 mL/min) c) Rinse with ultrapure water (25 min at 0.3 mL/min) d) Rinse with eluent (100 min at 0.3 mL/min)
Contamination by high-valency hydrophilic ions:
a) Rinse with ultrapure water (25 min at 0.3 mL/min) b) Rinse with 5% acetonitrile (20 min at 0.3 mL/min) c) Rinse with 100% acetonitrile (60 min at 0.3 mL/min) d) Rinse with 50% acetonitrile (10 min at 0.3 mL/min) e) Rinse with ultrapure water (50 min at 0.3 mL/min) f) Rinse with eluent (100 min at 0.3 mL/min)
System peak shifted (regeneration method with column oven):
a) Rinse with concentrated eluent 1 mol/L (25 min at 0.4 mL/min) b) Keep at 45 ... 50 °C for 10 ... 12 hours (no rinsing) c) Rinse with standard eluent (minimum 40 min at 0.4 mL/min) |